CPC A61B 5/1495 (2013.01) [A61B 5/0537 (2013.01); A61B 5/0538 (2013.01); A61B 5/14503 (2013.01); A61B 5/14532 (2013.01); A61B 5/1459 (2013.01); A61B 5/1473 (2013.01); A61B 5/14865 (2013.01); A61B 5/4839 (2013.01); A61B 5/6849 (2013.01); A61B 5/6852 (2013.01); A61B 5/6886 (2013.01); A61B 5/7203 (2013.01); A61B 5/7221 (2013.01); A61B 5/7225 (2013.01); A61B 5/7242 (2013.01); A61B 5/746 (2013.01); A61M 5/14276 (2013.01); A61M 5/1582 (2013.01); A61M 5/1723 (2013.01); G01N 27/026 (2013.01); G01N 27/028 (2013.01); G01N 27/416 (2013.01); G01N 27/4163 (2013.01); G01N 33/49 (2013.01); G01N 33/66 (2013.01); G01N 33/96 (2013.01); G01R 35/00 (2013.01); G01R 35/005 (2013.01); G08B 21/182 (2013.01); A61B 2562/0214 (2013.01); A61B 2562/04 (2013.01); A61M 5/14244 (2013.01); A61M 2005/1726 (2013.01)] | 20 Claims |
1. A method of calibration of a glucose sensor involving low start-up detection, the method comprising:
performing electrochemical impedance spectroscopy (EIS) procedures for a working electrode of the glucose sensor to obtain values of at least one impedance-based parameter for the working electrode;
performing a bound check for a Nyquist slope and a real impedance value at 1 kHz based on the at least one impedance-based parameter to calculate a reliability index;
performing a first calibration of the glucose sensor in a case where the reliability index is higher than a predetermined threshold; and
in a case where the reliability index is not higher than the predetermined threshold:
calculating a length of time for the Nyquist slope and the real impedance value at 1 kHz to be within a respective range;
performing the first calibration of the glucose sensor in a case where the length of time is shorter than a predetermined duration; and
performing a second calibration with an EIS-based calibration factor adjustment in a case where the length of time is not shorter than the predetermined duration.
|