CPC F16L 3/10 (2013.01) [A61M 5/00 (2013.01); A61M 5/142 (2013.01); A61M 5/14228 (2013.01); A61M 5/365 (2013.01); A61M 39/281 (2013.01); A61M 39/284 (2013.01)] | 16 Claims |
1. A lever, comprising:
a far end configured to engage an infusion tubing;
a pivot end configured to fit onto a shaft and to allow the lever to rotate about a shaft axis, the pivot end comprising a pocket to capture a spring having a leg protruding outside of the pocket, wherein the spring is configured to bias the lever against an infusion pump door assembly;
a tab disposed on an outer side of the far end and configured to interface with the door assembly; and
a pusher protruding into an inner side of the far end and configured to contact the infusion tubing when the far end engages the infusion tubing, wherein
the pusher is configured to apply a downward pressure to the infusion tubing when the door assembly is closed until the infusion tubing is properly aligned with an air-in-line sensor disposed on both sides of the infusion tubing,
wherein the far end comprises two walls projecting downward to engage the infusion tubing.
|